This is a sample notification!

الجهات المعتمدة
الحالة
تاريخ الانتهاء
رقم الشهادة
الاسم
نشاط المخبر
العنوان
البريد الالكتروني
الهاتف
الموقع الالكتروني
http://www.motabaqah.com
المتغيّر/الكمية المقيسة
المدى
أفضل إمكانية قياس
وصف مختصر للقياسات والمعدات المستخدمة
DC Voltage Measurement 100 uV to 1000 uV (1m to 100m)V (100m to 1000)V(33000 to 36) ppm (24 to 3)ppm (3 to 2.5)ppm Fluke 732B, 752A, 720A & Agilent 34420A Direct / Comparison
AC Voltage Measurement @ (10 to 20)Hz(2 to 200)mV, (200m to 200)V(2560 to 94)ppm (280 to 71)ppmFluke 792A, 5720A, 5808A Direct/Comparison
AC Voltage Measurement @ (20 to 1k)Hz(2 to 200)mV (200m to 200)V (200 to 1000)V(2560 to 47)ppm (94 to 34)ppm (71 to 34)ppmFluke 792A, 5720A, 5808A Direct/Comparison
AC Voltage Measurement @ (1k to 20k)Hz(2 to 200)mV (200m to 200)V (200 to 1000)V(1870 to 47)ppm (47 to 34)ppm 34ppmFluke 792A, 5720A, 5808A Direct/Comparison
AC Voltage Measurement @ (20k to 100k)Hz(2 to 200)mV (200m to 20)V(2211 to 700)ppm (20 to 556)ppmFluke 792A, 5720A, 5808A Direct/Comparison
AC Voltage Measurement @ (100k to 1M)Hz(2 to 200)mV (200m to 20)V(2211 to 700)ppm (20 to 556)ppmFluke 792A, 5720A, 5808A Direct/Comparison
DC Current Measurement (1u to 100u)A (100u to 1m)A (1m to 100m)A (100m to 10)A (10 to 20)A (20 to 300)A (135 to 8) ppm (8 to 7) ppm 7 ppm (7 to 9)ppm (9 to 22)ppm (22 to 62)ppmIET SRL Resistors,Guildline 9230A Shunt, Fluke 732B, 752A, 5720A, Agilent 34420A Direct/Comparison
AC Current Measurement @ (10 to 1k)Hz (1u to 1m)A (1m to 20)A (200 to 80)ppm 60ppm "Fluke 5790A & Metal Film Resistor Direct/Comparison Fluke A40A, 792A, 5808A,5720A Direct/Comparison
AC Current Measurement @ (1k to 20k)Hz (1m to 1)A (1 to 20)A 60 ppm (235 to 243)ppm"Fluke 5790A & Metal Film Resistor Direct/Comparison Fluke A40A, 792A, 5808A,5720A Direct/Comparison
DC Resistance (1 m to 1)Ω (1 to 10)Ω (10 to 10k)Ω (10k to 10M)Ω (10M to 1G)Ω(67 to 11)ppm (11 to 7.5)ppm (7.5 to 4.5)ppm (4.5 to 9.5)ppm (9.5 to 495)ppmIET SRL, Fluke 8508A Direct/Comparison
Phase Angle Measurement (10 to 1k)Hz (1k to 30k)Hz ±180˚ ±180˚ 0.26˚ 1.8˚ "Krohn-Hite 6610 Direct/Comparison"
Temperature Measuremetn (Electrical Measurement Type - K Type - J Type - E Type - T Type - N Type - R Type - S Type - B Type - C RTD(-200 to 1350)ºC (-200 to 1200)ºC (-250 to 1000)ºC (-250 to 400)ºC (-250 to 1300)ºC (0 to 1750)ºC (0 to 1750)ºC (600 to 1800)ºC (0 to 2300)ºC (-200 to 800)ºC(0.0093 to 0.014)ºC (0.001 to 0.0098)ºC (0.017 to 0.008)ºC (0.021 to 0.004)ºC (0.001 to 0.012)ºC (0.023 to 0.019)ºC 0.023 ºC (0.021 to 0.018)ºC (0.009 to 0.038)ºC (0.001 to 0.014)ºCFluke 8508A
DC Resistance Source Fixed 0.001Ω 0.01 Ω 0.1 Ω 1 Ω 10 Ω 100 Ω 1 kΩ 10 kΩ 100 kΩ 1 MΩ 10MΩ 100MΩ 1000MΩ126 60 ppm 116 ppm 461 ppm 20 ppm 10.6 ppm 8.3 ppm 7.5 ppm 10.4 ppm 7.4 ppm 10 ppm 12.7 ppm 40 ppm 180 ppmIET SRL, Fluke 742A, IET SRL, Fluke 742A, IET SRL, Fluke 8508A-7000k
DC Resistance Source Continious 1 to 10 Ω 10 to 30 Ω 30 to 100 Ω 100 to 300 Ω 0.0003 to 1 MΩ 1 to 3 MΩ 3 to 10 MΩ 10 to 30 MΩ 30 to 100 MΩ 100 to 300 MΩ 300 to 1100 MΩ(0.12 to 0.017) % (0.017 to 0.0097) % (0.0097 to 0.0051) % (0.0051 to 0.0042) % 0.0042 % (0.0042 to 0.0085) % (0.0085 to 0.016) % (0.016 to 0.039) % (0.039 to 0.063) % (0.063 to 0.39) % (0.39 to 1.8) %Fluke 5520A
AC Resistance Source @ (0.1M to 1M)Hz 10kΩ 100kΩ0.08% (0.08 to 0.12)%Agilent 42030A Direct
AC Resistance Source @ (1M to 13M)Hz 10Ω 100Ω 1kΩ(0.12 to 1.5)% (0.12 to 0.7)% (0.13 to 0.6)%Agilent 42030A Direct
Capacitance Source (120 to 1k)Hz (1k to 10k)Hz (10k to 100k)Hz (100k to 1M)Hz (1M to 13M)Hz(10n,100n,1u)F (10p,100p,1n,10n,100n,1u)F (1p,10p,100p,1n,10n,100n,1u)F (1p,10p,100p,1n)F (1p,10p,100p,1n)F 0.12% (0.12 to 0.14)% (0.16 to 0.14)% 0.16% (0.16 to 2.5)%Agilent 16380A, 16380C
AC Power @ (45 to 65)Hz (1.1m to 3m)W (3m to 20k)W(0.55 to 0.16)% (0.16 to 0.12)%Fluke 5520A
Oscilloscope DC Amplitude @ 50Ω (1m to 10m)V (10m to 6.6)V @ 1kHz, Square Wave (1m to 10m)Vp-p (10m to 6.6)Vp-p Square Wave 1MΩ (1m to 10m)Vp-p (10m to 6.6)Vp-p Time Base (500p to 2n)s (2n to 5)s Band Width (50k to 250k)Hz (250k to 6G)Hz (3 to 0.4)% (0.4 to 0.04)% (5 to 0.8)% (0.8 to 0.3)% (0.7 to 0.15)% (0.15 to 0.1)% 0.00% (0.0015 to 0.0006)% 4% 4%Fluke 5820A E4438C, Fluke 910R Fluke 5820A Fluke 5820A, 910R,Agilent E9304A&E4419B Fluke 910R,Agilent E4438C,E9304A&E4419B
Temperature Source (Electrical Simulation Type - K Type - J Type - E Type - T Type - N Type - R Type - S Type - B Type - C RTD, PT-385, 100Ω (-200 to 1350)ºC (-200 to 1200)ºC (-250 to 1000)ºC (-250 to 400)ºC (-250 to 1300)ºC (0 to 1750)ºC (0 to 1750)ºC (600 to 1800)ºC (0 to 2300)ºC (-200 to 800)ºC (0.032 to 0.028)ºC (0.003 to 0.019)ºC (0.055 to 0.015)ºC (0.074 to 0.011)ºC (0.001 to 0.025)ºC (0.093 to 0.05)ºC (0.093 to 0.06)ºC (0.08 to 0.053)ºC (0.036 to 0.088)ºC (0.001 to 0.014)ºCFluke 5720A Fluke 8508A & DRB
RF & Microwave Frequency Measurement (10 to 12.4G)Hz (12.4G to 46G)Hz (46G to 50G)Hz (6µ to 1.2)Hz (1.2 to 2.1)Hz 2.1HzFluke 910R, 53132A 53152A E4448A
Power Measurement @ (100k to 10M)Hz (-35 to 20) dBm (20 to 44) dBm (0.09 to 0.06)dB (0.06 to 0.26)dBN8482A,8482B
Power Measurement @ (10M to 4.2G)Hz (-70 to -30) dBm (-30 to -20) dBm (-20 to 10) dBm (10 to 20) dBm (20 to 44) dBm (0.38 to 0.13)dB (0.13 to 0.09)dB (0.04)dB (0.06)dB (0.06 to 0.26)dB8481D N8485A 8478b,432A N8482A 8482B
Power Measurement @ (4.2G to 18G)Hz (-70 to -30) dBm (-30 to -20) dBm (-20 to 10) dBm (10 to 20) dBm(0.38 to 0.19)dB (0.13 to 0.09)dB (0.13 to 0.09)dB (0.07 to 0.08)dB 8481D N8485A 8478B,432A N8481A
Power Measurement @ (18G to 50G) Hz(-70 to -30) dBm (-30 to 20) dBm(0.38 to 0.24)dB (0.11 to 0.17)dB 8487D N8487A
Modulation Measurement Amplitude Modulation Carrier Frequency= 150kHz to 1.3GHz Rate= 20Hz to 100kHz Modulation Depth 5 to 99 % (1.4 to 3.6)%" 8902A
Frequency Modulation Carrier Frequency= 150kHz to 1.3GHz Rate = 20Hz to 200kHz Freq.Deviation 10kHz to 400kHz (1.4 to 6.6)% 8902A
Phase Modulation Carrier Frequency= 150kHz to 1.3GHz Rate = 200Hz to 20kHz Phase Deviation 10° to 400° 4.80% 8902A
Distortion 20Hz to 100kHz THD (-99.99 to 0)dB (1.2 to 2.4)dB 8903B
Digital Modulation Magnitude EVM for BPSK, QPSK, 8PSK, DQPSK,D8PSK, π /4 DQPSK Phase EVM for BPSK, QPSK, 8PSK, DQPSK,D8PSK, π /4 DQPSK Symbol rate < 1MHz < 6MHz Symbol rate < 1MHz < 6MHz 1 %rms 2.5%rms 0.6 %rms 1.5%rms E4440A
Phase Noise Carrier Frequency= (50k to 26.5G)Hz Offset Frequency (0.01 to 100 M)Hz 4 dB E4440A
Reflection Coefficient (9k to 50G)Hz Magnitude Phase 0 to 1 ±180° 0.004 to 0.06 E5071C, N5230C
Frequency Source (10 to 12.4G)Hz (12.4G to 46G)Hz (46G to 50G)Hz (6µ to 1.2)Hz (1.2 to 2.1)Hz 2.1Hz Fluke 910R E8247C, E4431B
Attenuation Source (50M to 18G) Hz (18G to 26.5G) Hz (1 to 110)dB (10 to 70)dB (0.36 to 2.4)dB (0.48 to 3.1)dB 8494H, 8496H 8495D
Modulation Generate Amplitude Modulation Carrier Frequency= (11M to 13.5M)Hz Rate= (20 to 100k)Hz Modulation Depth (5 to 99) % 0.4 % rdg 11715A
Frequency Modulation Carrier Frequency= (11M to 432M)Hz Rate = (100k to 200k)Hz Freq. Deviation (10k to 400k)Hz (10k to 400k)Hz 0.40% 3.00% 11715A
Phase Modulation Carrier Frequency= (11M to 432M)Hz Rate= (20 to 20k)Hz Phase Deviation (0.1 to 400)° (1 to 3)% 11715A
Mass (Measure & Generate) 1 mg 2 mg 5 mg 10 mg 20 mg 50 mg 100 mg 200 mg 500 mg 1 g 2 g 5 g 10 g 20 g 50 g 100 g 200 g 500 g 1 kg 2 kg 5 kg 10 kg 20 kg 50 kg 500 kg 0.0071 mg 0.0066 mg 0.0066 mg 0.0066 mg 0.0066 mg 0.0071 mg 0.0077 mg 0.0084 mg 0.0099 mg 0.025 mg 0.026 mg 0.028 mg 0.04 mg 0.057 mg 0.086 mg 0.18 mg 0.38 mg 0.85 mg 0.0018 g 0.0036 g 0.0085 g 0.019 g 0.037 g 0.096 g 32.1 g OIML Ultra Class Weight OIML E2 Class Weight OIML M1 Class Weight
Torque (5 to 1000) lb.in (25 to 2000) lb.ft 0.30% 0.30% CDI
Pressure Differential Pressure (0 to 1) psi ( 0.0023) psi Fluke 525A
Absolute Pressure ( 0 to 15) psi ( 15 to 1000) psi (0.00085 to 0.0021) psi (0.0021 to 0.13) psi Fluke PPC3
Gauge Pressure (-14.5 to 15) psi (15 to 750) psi (750 to 2000) psi (2000 to 75000) psi 0.0021 psi 0.0021 to 0.13 psi (0.011 to 0.0092) % (0.0092 to 0.0082) % Fluke PPC3 CPB6000 HXS DWT
Sound Level Meter 31.5Hz to 8 kHz @12.5kHz @16kHz 31.5Hz to 8 kHz @12.5kHz @16kHz 31.5Hz to 8 kHz @12.5kHz @16kHz 94 dB 94 dB 94 dB 104 dB 104 dB 104 dB 114 dB 114 dB 114 dB 0.25 dB 0.28 dB 0.31 dB 0.36 dB 0.38 dB 0.40 dB 0.36 dB 0.38 dB 0.40 dB B&K 4226
Temperature (Measure & Source) (-80 to 0) °C (0 to 400) °C (400 to 660) °C (660 to 1100) °C 0.013 °C 0.029 °C 0.05 °C 1.61 °C Fluke 5628 & 1529A Fluke 5650 & 1529
Humidity 0.21 % RH 9.96 % RH 34.98 % RH 50.1 % RH 79.94 % RH 0.3 % RH 0.3 % RH 0.5 % RH 0.9 % RH 1.2 % RH Rotronic Humidity Standard
IR Thermometer (-15 to 0) °C (0 to 500) °C (500 to 1100) °C 0.14 °C 0.53 °C 1.61 °C Fluke 5618B, 1529 Fluke 5650, 1529
Pipette(0.001 to 20) ml 0.58 µl XP26PC
Volume (20 to 200) ml (0.2 to 10) l 6.1 µl 0.58 ml Weighing Balance
Angle Generate (1 to 90) ° 25 arcs AB-10-ST
Vernier Caliper & Height Gauge 0 to 300 mm 300 to 600 mm 600 to 1000 mm 5.8 to 6.0 µm 6.0 to 6.5 µm 6.5 to 7.5 µm Mahr 415/1
Micrometer 0 to 50 mm 0.6 µm Mahr 409C/0
Dial Indicator 0 to 25 mm 3 µm Aditya SYLVAC D50
Scale & Tape 0 to 1000 mm 0.29 mm Octogan MSTC 1000
Gauge Block 0 to 100 mm 0.15 to 0.35 µm Mahr 409C/0, Millimar 1240/826
Length & Thicknes Measuring Tools 0 to 100 mm 0.68 µm Mahr 409C/0,