الكمية المقيسة إمكانية المعايرة والقياس الوصف المختصر للمعدات المدى
Acoustic
Sound Level Meters 0.9 dB Calibration of sound level meters Equipment Used: Acoustic calibrator Calibration Procedure LABC-020 (ANSI S1.4) 94 to 114 dB @ 12 kHz
Capacitance
Capacitance Generate 16 pF 0.023 nF 0.22 nF 0.22 nF 0.28 nF 2.20 nF 5.30 nF 15.30 nF 19 nF 0.045 μF 0.029 μF 0.15 μF 0.24 μF 2.0 μF 1.5 mF 0.070 mF 0.32 mF 1.49 Mf Capacitance Generate Equipment Used: Multi-product calibrator Calibration Procedure LABC-018 (EURAMET cg-15) (10 Hz to 10 kHz) 220.0 pF to 399.9 pF 0.4 nF to 1.0999 nF (10 Hz to 3 kHz) 1.16 nF to 3.2999 nF (10 Hz to 1 kHz) 3.3 nF to 10.9999 nF 11 nF to 32.999 nF 32.999 nF 109.999 nF 110 nF to 329.999 nF (10 Hz to 600 Hz) 0.33 μF to 1.09999 μF (10 Hz to 300 Hz) 1
Capacitance Measure 0.053 nF 0.17 nF 1.7 nF 18 nF 0.17 μF 0.35 μF 18 μF 0.17 mF 4.8 mF Capacitance Measure Equipment Used: Precision multimeter Calibration Procedure LABC-018 (EURAMET cg-15) 0 nF to 1 nF 1 nF to 10 nF 10 nF to 100 nF 100 nF to 1 μF 1 μF to 10 μF 10 μF to 100 μF 100 μF to 1 mF 1 mF to 10 mF 10 mF to 100 mF
Chemical
Dissolved oxygen meters. 0.012 Dissolved oxygen meters. Equipment used: Standard Solutions Calibration Procedure LABC-089 (ASTM D888) Zero Calibration
PH 0.06 pH Calibration of PH Meters Equipment used: Standard Solutions Calibration Procedure LABC-074 (ASTM D1293) 0 pH to 14 pH
TDS 0.91 ppm 7.2 ppm 21 ppm Calibration of TDS Meters Equipment used: Standard Solutions Calibration Procedure LABC-026 (ASTM D5907) 0 ppm to 100 ppm 100 ppm to 703ppm 703 ppm to 2500 ppm
Viscometer 0.011 Calibration of viscometer, Equipment Used: standard solution Calibration Procedure LABC-057 (ASTM D445) 500 CPS to 5000 CPS
Gas Detectors 0.02 Calibration of gas detectors Equipment used: Span calibration gases Calibration Procedure LABC-023 (BS EN 60079-29) Methane: 50% LEL, O2: 20.9% H2S: 25ppm CO: 100ppm
Conductivity
Conductivity Meters 0.08 µs/cm 0.13 µs/cm 3.9 µs/cm 6.9 µs/cm 6.9 µs/cm 14. µs/cm Calibration of Conductivity Meters Equipment Used: Standard Solutions Calibration Procedure LABC-047 (ASTM D1125) 0 µs/cm to 5 µs/cm 5 µs/cm to 10 µs/cm 10 µs/cm to 250 µs/cm 250 µs/cm to 500 µs/cm 500 µs/cm to 950 µs/cm 950 µs/cm to 1413 µs/cm
Current
Level Transmitters 0.013 mAdc Level Transmitters and switches (onfield verification) Equipment Used: Process Calibrator Calibration Procedure LABC-112 (ISO 4266-1, BS EN 13616-2, ASTM D5413) 0 mAdc to 20 mAdc