Details
of Accredited Bodies
Measured Quantity Calibration and Measurement Capability (CMC) Brief Description Of Measurement Range For Instrument
Capacitance
Capacitance Measure 0.59% 0.41% 0.67% Comparasion with Fluke 5522A Euramet CG-15 up to 1 nF 1 nF to 10.9 uF 10.9 uF to 109 uF
Capacitance Source 0.27 % 0.13 % Direct comparasion with Fluke 8588A LMR-WI-TCH-001 up to 1 nF 1 nF to 200 uF
Current
AC Current Measure @ 1 kHz 0.13% 0.059% 0.19% 0.11% Comparasion with Fluke 5522A Euramet CG-15 up to to 33 mA 33 mA to 329 mA 329 mA to 2.99 A 2.99 A to 10 A
DC Current Mreasure 0.013 % 0.027 % 0.064 % 0.045 % Comparasion with Fluke 5522A Euramet CG-15 up to 329 mA 329 mA to 1.09 A 2.99 A to 10 A 1.09 A to 2.99 A
AC Current Source @ 1 kHz 0.15 % 0.032 % 0.11 % 0.058 % Direct comparasion with Fluke 8588A LMR-WI-TCH-001 up to 10 mA 10 mA to 100 mA 1 A to 20 A 100 mA to 1 A
DC Current Source 0.0023 % 0.0078 % 0.025 % Direct comparasion with Fluke 8588A LMR-WI-TCH-001 up to 10 mA 10 mA to 100 mA 100 mA to 20 A
Mass
Truck Scale 1.4 kg +0.6*Resolution 2.9 kg +0.6*Resolution Standard Mass class M3; OIML R76-1 500 to 10 000 kg 10 000 to 20 000 kg
Balance 0.43 mg +0.6*Resolution 5.8 mg +0.6*Resolution 4.0 g +0.6*Resolution 1.4 g +0.6*Resolution Standard Mass class E2, M3; OIML R76-1 up to 200 g > 220 g to 5 kg >25 kg to 300 kg > 5 kg to 25 kg
Resistance
Resistance Measure 0.0049 % 0.0035 % 0.0051 % 0.0063 % 0.0082 % 0.0058 % 0.0081 % 0.0046 % 0.0071 % 0.0048 % 0.0094 % 0.0071 % 0.015 % 0.042 % 0.033 % 0.073 % 0.35 % 1.8 % Comparasion with Fluke 5522A Euramet CG-15 up to 10 Ω 10 Ω to 1.9 kΩ 1.9 kΩ to 3 kΩ 3 kΩ to 3.3 kΩ 3.3 kΩ to 10.9 kΩ 10.9 kΩ to 11.9 kΩ 11.9 kΩ to 19 kΩ 19 kΩ to 30 kΩ 30 kΩ to 33 kΩ 33 kΩ to 1 MΩ 1 MΩ to 1.9 MΩ 1.9 MΩ to 3 MΩ 3 MΩ to 10.9 MΩ 10.9 MΩ to 11.9 MΩ 11.9 MΩ to 30 MΩ 30 MΩ to 109 MΩ 109 MΩ to 290 MΩ 290 MΩ to 1090 MΩ
Resistance Source 0.0022 % 0.00089 % 0.00045 % 0.00029 % 0.00091 % 0.012 % 0.026 % Direct comparasion with Fluke 8588A LMR-WI-TCH-001 up to 10 Ω 10 Ω to 100 Ω 100 Ω to 1 kΩ 1 kΩ to 100 kΩ 100 kΩ to 10 MΩ 10 MΩ to 100 MΩ 100 MΩ to 1 GΩ
Sec
Active: The accreditation is valid.
Sec
Expired: Accreditation has expired and, in this case, it is considered withdrawn.
Sec
Cancelled: Accreditation has been cancelled based on the body’s request or the center’s decision, and in this case, it is considered withdrawn.
Sec
Suspended: Accreditation has been suspended at the body's request or based on SAAC's decision